• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) AVEIR¿ LEADLESS PACEMAKER, RIGHT VENTRICLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) AVEIR¿ LEADLESS PACEMAKER, RIGHT VENTRICLE Back to Search Results
Model Number LSP112V
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2024
Event Type  malfunction  
Event Description
It was reported the patient presented for implant procedure.During surgery, the leadless pacemaker exhibited a failure to interrogate over conductive telemetry.The surgery was completed with a different leadless pacemaker.The patient was in stable condition.
 
Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.The reported event of failure to interrogate was confirmed.The device was unable to establish telemetry and no output was noted upon receipt.Analysis after the device was cut open revealed battery voltage was above elective replacement indicator (eri) level, which should have been capable to support telemetry.Further analysis performed on the hybrid indicated a metal migration anomaly causing a short.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVEIR¿ LEADLESS PACEMAKER, RIGHT VENTRICLE
Type of Device
LEADLESS PACEMAKER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18576987
MDR Text Key333698780
Report Number2017865-2024-01974
Device Sequence Number1
Product Code PNJ
UDI-Device Identifier05415067034472
UDI-Public05415067034472
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P150035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLSP112V
Device Lot NumberS000091991
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age86 YR
Patient SexMale
Patient Weight77 KG
-
-