NIHON KOHDEN CORPORATION GZ-130PA; TRANSMITTER Back to Search Results
Model Number GZ-130PA
Device Problem Patient Data Problem (3197)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/02/2024
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer (bme) reported that the gz transmitter will randomly show flat line for all ecg waves.The bme tried different cables and ran the unit on a simulator and has reproduced the issue.There are no waves on the unit itself.No patient harm was reported.No physical damage to the device.Biomed requested a warranty exchange on the device and will return this unit.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The biomedical engineer (bme) reported that the gz transmitter will randomly show flat line for all ecg waves.The bme tried different cables and ran the unit on a simulator and has reproduced the issue.There are no waves on the unit itself.No patient harm was reported.No physical damage to the device.Biomed requested a warranty exchange on the device and will return this unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GZ-130PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 116-8 560
JA  116-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key18577001
MDR Text Key334731785
Report Number8030229-2024-04055
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921117415
UDI-Public04931921117415
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153707
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGZ-130PA
Device Catalogue NumberGZ-130PA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/21/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-