It was reported that the patient was on support for 4-5 hours and then alarms were displayed, and the device had no flow.The failure occurred during treatment.The hls set was transferred to the emergency drive and flow could be achieved.They transferred the hls set to another cardio help and were able to achieve flow for 5-10 minutes and on the flow stopped again.The error message ¿pump disposable error¿ was displayed.The hls set was again transferred to the emergency drive and a new hls set was primed.After the hls set was exchanged, the failure did not reoccurred with the new hls set.No harm to any person has been reported.Complaint id (b)(4).
|