• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH TUBING SET; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH TUBING SET; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HLS SET
Device Problems Decoupling (1145); Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/21/2024
Event Type  malfunction  
Event Description
It was reported that the patient was on support for 4-5 hours and then alarms were displayed, and the device had no flow.The failure occurred during treatment.The hls set was transferred to the emergency drive and flow could be achieved.They transferred the hls set to another cardio help and were able to achieve flow for 5-10 minutes and on the flow stopped again.The error message ¿pump disposable error¿ was displayed.The hls set was again transferred to the emergency drive and a new hls set was primed.After the hls set was exchanged, the failure did not reoccurred with the new hls set.No harm to any person has been reported.Complaint id (b)(4).
 
Manufacturer Narrative
A follow-up medwatch will be submitted when additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUBING SET
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18579953
MDR Text Key333776983
Report Number8010762-2024-00049
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112360
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHLS SET
Device Catalogue Number701069078
Device Lot Number3000342903
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-