Event Description
It was reported a mitraclip procedure was performed to treat functional mitral regurgitation (mr) and tricuspid regurgitation (tr) with a grade of 5+.A steerable guide catheter (sgc) was inserted and three clips were successfully deployed on the mitral valve without issues.The sgc was then retracted into the right atrium (ra) and one mitraclip was deployed on the tricuspid valve.However, after deployment, it was suspected the clip detached from the anterior leaflet and remained attached to the septal leaflet (single leaflet device attachment/slda).Tr was reduced to a grade of 3-4 and no additional treatment was performed.After removal of the sgc, a left to right shunt was observed.Therefore, an atrial sept defect (asd) closure device was implanted.No clinically significant delay in the procedure.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the cause of the reported slda was unable to be determined.The reported off-label use was associated with the use of the mitraclip device on the tricuspid valve.There is no indication of a product quality issue with respect to manufacture, design, or labeling the additional mitraclip device referenced in b5 is being filed under a separate medwatch report.H6: device code 1494 - indication for use (use in tricuspid valve).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL NORTHPOINT REG 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18580212
MDR Text Key334065844
Report Number2135147-2024-00352
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037381
UDI-Public05415067037381
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0706-XTW
Device Lot Number31003A2065
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Age88 YR
Patient SexFemale
Patient Weight65 KG
-
-