It was reported that the web device was used in a patient to treat an unruptured aneurysm in the middle cerebral artery (mca).When the web was deployed in position it did not appear to detach when attempted multiple times using the detachment controller.When the pusher was pulled back to check if the web was detached or not, the web came out from the aneurysm.During re-sheathing attempt, the web unintentionally detached outside the aneurysm.The physician used a snare to capture, re-sheath and remove the web from the patient.A new web and via-21 catheter were used to continue the procedure.Reportedly, three attempts were made to detach the second web with a controller, but the web could not be detached.The physician advanced the microcatheter to contact the web and mechanically detached the web in the aneurysm and concluded the procedure.There was no report of harm or injury to the patient.This is report 1 of 2 for this event.
|
Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging was not provided for this investigation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): please refer to the japanese ifu for precautions, warnings, and further information.The following is taken from the english version: potential complications potential complications include but are not limited to the following: hematoma at the site of entry, aneurysm rupture, emboli, vessel perforation, parent artery occlusion, hemorrhage, ischemia, vasospasm, clot formation, device migration or misplacement, premature or difficult device detachment, non-detachment, incomplete aneurysm filling, revascularization, post-embolization syndrome, and neurological deficits including stroke and death.Warnings and precautions the web aneurysm embolization system is intended for single use only.The detachment control device is intended to be used for one patient.Do not resterilize and/or reuse the device.Reuse and/or resterilization can increase risk of infection, cause a pyrogenic response or other life threatening complications.Reuse and/or resterilization can degrade product performance, leading to device malfunction.Dispose of all devices in accordance with applicable hospital, administrative and/or local government policy.Advance and retract the device slowly.Do not advance the delivery device with excessive force.Determine the cause of any unusual resistance.Remove the device if excessive friction is noted and check for damage.Do not rotate the delivery device during or after delivery of the embolization device.Rotating the device may result in damage or premature detachment.The embolization device cannot be detached with any other power source other than a microvention inc.Detachment control device.Ensure that at least two detachment control devices are available before initiating an embolization procedure.Procedure detachment of the device 31.The detachment control device is pre-loaded with batteries and will activate when the delivery device is properly connected.32.Verify that the rhv is firmly locked around the delivery device before attaching the detachment control device to ensure that the embolization device does not move during the connection process.33.Ensure that the delivery device gold connectors are clean and free from blood or contrast.If necessary, wipe the connectors with sterile water and dry before connecting.34.Insert the proximal end of the delivery device into the detachment control device.When the delivery device is properly connected, the light will flash green and an intermittent tone will be heard.35.Verify the embolization device position before pressing the detachment button.36.Push the detachment button.During firing, the light should be solid green and the beep should be continuous.37.Verify detachment by first loosening the rhv valve, then pulling back slowly on the delivery device and verifying that there is not embolization device movement.If the embolization device does not detach, push the detachment button again.If the device is still not detached, obtain a new detachment control device and attempt detachment up to two additional times.If it does not detach, remove the delivery device.38.Verify the position of the embolization device angiographically through the guide catheter.39.Prior to removing the microcatheter from the treatment site, place an appropriately sized guidewire completely through the microcatheter lumen to ensure that no part of the embolization device remains within the microcatheter.Investigation conclusion: the physical device was not available for evaluation to determine if a condition existed that would have caused or contributed to event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.This information may be updated if additional information is provided at a later date.
|