MEDTRONIC EUROPE SARL COBALT¿ HF QUAD CRT-D MRI SURESCAN¿; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTPB2QQ |
Device Problems
Electromagnetic Interference (1194); Mechanical Problem (1384); Unable to Obtain Readings (1516)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/10/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during use of the right atrial (ra) lead there was a single atrial tachycardia/atrial fibrillation (af) episode caused by noise. pocket manipulations were performed and atrial oversensing was noted. the ra lead remains in use. noted that the cardiac resynchronization therapy defibrillator (crt-d) may be incrementing af evidence, but episode criteria may not be met. it was also reported that the patient was seen by the clinic for evaluation and 60hz "poss" overspending was seen and suspected possible setscrew issue.The crt-d remains in use.It was reported that the remote monitoring network experienced invalid data.Clinical data was reviewed.Optivol data as "invalid" in the quick look observation.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated noise.Correction: b5 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during use of the right atrial (ra) lead there was a single atrial tachycardia/atrial fibrillation (af) episode caused by noise. pocket manipulations were performed and atrial oversensing was noted. the ra lead remains in use. noted that the cardiac resynchronization therapy defibrillator (crt-d) may be incrementing af evidence, but episode criteria may not be met. it was also reported that the patient was seen by the clinic for evaluation and 60hz possible oversensing was seen and suspected possible setscrew issue.The crt-d remains in use.It was also reported that the remote monitoring network report contained invalid data.Diagnostic interpretation was provided.No further patient complications have been reported as a result of this event.
|
|