ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
An investigation into the reported event has been initiated under (b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.
 
Event Description
It was reported that the unit was missing two screws on the bottom.There is no information provided regarding the timing of the event or if there was any patient involvement/harm that may have occurred.Due diligence is in progress, no further information is available.
 
Event Description
There was no additional information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).The following sections have been corrected/updated: b4, b5, d2, d4, d9, g1, g3, g6, h1, h2, h3, h4, h6, and h10.Review of the most recent repair record determined various components were damaged and corroded.The motor, sleeve, spring seal, swivel, lever, control bar, width plate, bearings, screws, pins, planetary carrier, o-ring, and various other parts were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18586228
MDR Text Key333836940
Report Number0001526350-2024-00137
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375901
UDI-Public(01)00889024375901(11)220314(10)65438335
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number65438335
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-