This event has been recorded by zimmer biomet under (b)(4).The following sections have been corrected/updated: b4, b5, d2, d4, d9, g1, g3, g6, h1, h2, h3, h4, h6, and h10.Review of the most recent repair record determined various components were damaged and corroded.The motor, sleeve, spring seal, swivel, lever, control bar, width plate, bearings, screws, pins, planetary carrier, o-ring, and various other parts were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|