The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured on (b)(6), 2018.One decontaminated sample was received for evaluation.A visual inspection was performed according to procedure.Additionally, a functional test was performed.The leak test was performed by inflating the balloon and the reported condition was observed; a leakage was observed at the bottom of the tube, confirming the reported condition.Based on historical data the supplier was issued a formal corrective/preventative action (capa) the root cause and action plan will be documented to address the reported condition to mitigate any further occurrences.This action is currently ongoing.This complaint will be closed with no further action and will be used for tracking and trending purposes.
|