Model Number 3851
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2023
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on 04jan2024.It was reported that a balloon leak occurred.A 10mmx2.75mm wolverine coronary cutting balloon was selected for use.During procedure, the cutting balloon leaked gas and could not be used; but there was no pinhole noted.The procedure was completed with another of the same device.No patient complications were reported.However, device analysis revealed a pinhole leak 4mm distal from proximal markerband.
 
Manufacturer Narrative
E1: initial reporter address 1: (b)(6).The device was returned to the for analysis.Visual, tactile and microscopic analysis was performed on the device.Examination of the device identified a pinhole and a kink along the hypotube shaft.A hypotube kink was noted 82.5cm distal to the distal end of the strain relief.A detailed microscopic examination of the balloon material identified a pinhole leak 4mm distal from proximal markerband when pressure was applied.No other device issues were identified during returned product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18589382
MDR Text Key334842293
Report Number2124215-2024-03955
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0031501126
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age82 YR
Patient SexMale
Patient Weight62 KG
-
-