Model Number 471205-17 |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/03/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) has not received the fenestrated bipolar forceps instrument for evaluation.Therefore, the root cause of the customer reported failure mode has not been determined.A follow-up mdr will be submitted if the product is returned and evaluated and/ or if additional information is received.
|
|
Event Description
|
It was reported that during central processing, the fenestrated bipolar forceps instrument ceramic in the mouth area was damaged by electricity.No known impact or patient consequence was reported.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the customer informed the instrument had been checked before use and no damage was observed.The customer noted that defective instruments are not reused.
|
|
Event Description
|
Refer to h10/h11 for follow-up information.
|
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The fenestrated bipolar forceps instrument was analyzed and found to have charring and localized melting at the distal clevis's grip base.The instrument passed the electrical continuity test.
|
|