|
Model Number AF1100MBP |
Device Problem
Pressure Problem (3012)
|
Event Date 06/30/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The manual resuscitator may be the source of abnormal fico2 elevation.This is the issue that was referred to in the field.Notification initiated on oct 11, 2023.In addition, the fda medwatch report was received.
|
|
Event Description
|
Patient presentation: an unresponsive patient, post cardiopulmonary arrest, was assessed with a glasgow coma scale (gcs) score of 3, indicative of severe consciousness impairment.Medical interventions: rapid sequence intubation (rsi) was performed to establish airway patency.The endotracheal tube (ett) was ventilated with an adult sunmed (af1100mb-p) manual resuscitator with a manometer, positive end-expiratory pressure (peep) valve set at 15 cmh2o, heat and moisture exchange filter (hmfe), in-line end-tidal co2 (etco2) sensor and connected to a 15 l/min oxygen source.Fic02 elevation issue: while using the sunmed manual resuscitator, the patient's fraction of inspired co2 (fico2) levels increased until the clinical team placed the patient on a ventilator.This increase began at 10:24.20 (figure 1) and peaked at 36 mmhg at 10:25.43 (figure 2).Fico2 levels fluctuated between 10 and 30 mmhg (figure 3) until the patient was transitioned to a ventilator at 10:26.51.The elevation in fico2 resolved once the ventilator was initiated (figure 4).Product problem: suspicions arise that the manual resuscitator may be the source of the abnormal fic02 elevation.The fic02 increase poses a risk of hypercapnia, which can significantly impact the patient's bodily functions, physical well-being, and potentially lead to life-threatening consequences.Furthermore, this inconsistency raises concerns regarding adherence to performance specifications and user expectations.Patient outcome: the patients outcome is unknown, as care was handed over to another provider.
|
|
Manufacturer Narrative
|
The manual resuscitator may be the source of abnormal fico2 elevation.This is the issue that was referred to in the field notification initiated on oct 11, 2023.In addition, the fda medwatch report was received.The complaint of "the manual resuscitator may be the source of abnormal fico2 elevation" regarding part af1100mbp was confirmed to be associated with ncmr-02506.A resolution letter was not requested by the customer.
|
|
Event Description
|
Patient presentation: an unresponsive patient, post cardiopulmonary arrest, was assessed with a glasgow coma scale (gcs) score of 3, indicative of severe consciousness impairment.Medical interventions: rapid sequence intubation (rsi) was performed to establish airway patency.The endotracheal tube (ett) was ventilated with an adult sunmed (af1100mb-p) manual resuscitator with a manometer, positive end-expiratory pressure (peep) valve set at 15 cmh2o, heat and moisture exchange filter (hmfe), in-line end-tidal co2 (etco2) sensor and connected to a 15 l/min oxygen source.Fic02 elevation issue: while using the sunmed manual resuscitator, the patient's fraction of inspired co2 (fico2) levels increased until the clinical team placed the patient on a ventilator.This increase began at 10:24.20 (figure 1) and peaked at 36 mmhg at 10:25.43 (figure 2).Fico2 levels fluctuated between 10 and 30 mmhg (figure 3) until the patient was transitioned to a ventilator at 10:26.51.The elevation in fico2 resolved once the ventilator was initiated (figure 4).Product problem: suspicions arise that the manual resuscitator may be the source of the abnormal fic02 elevation.The fic02 increase poses a risk of hypercapnia, which can significantly impact the patient's bodily functions, physical well-being, and potentially lead to life-threatening consequences.Furthermore, this inconsistency raises concerns regarding adherence to performance specifications and user expectations.Patient outcome: the patients outcome is unknown, as care was handed over to another provider.
|
|
Search Alerts/Recalls
|
|
|