• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-Q290V
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2024
Event Type  malfunction  
Event Description
The customer reported to olympus, the duodenovideoscope forceps elevator became tangled with the stent, making it impossible to move.Once the scope was removed, the tangles were removed and the procedure was completed with the same set of equipment.The issue was found during a therapeutic stenting procedure.There were no reports of patient harm.
 
Manufacturer Narrative
The device was not returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the customer reportable malfunction was most likely caused by the stress of repeated use, external factors, or handling.However, the root cause of the events could not be determined.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: the instruction manual operation manual ¿chapter 3 preparation and inspection, 3.3 inspection of the endoscope¿ describes the methods for inspection on the suggested event.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18596889
MDR Text Key334781918
Report Number9610595-2024-01846
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-Q290V
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
STENT.
-
-