It was reported by japan that during service and evaluation, it was determined that the small battery drive device had moving parts that did not move smoothly-trigger and had insufficient/low power.It was further determined that the device failed pretest for check for sticky triggers and check power with power test bench.It was noted in the service order that the device had low power along with a small battery drive casing device.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.D1: concomitant med products and therapy dates: small battery drive casing device, 1/16/2024
device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition that the device had moving parts that did not move smoothly-trigger, identified during service and repair, was confirmed.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
|