• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST; MARYLAND BIPOLAR FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST; MARYLAND BIPOLAR FORCEPS Back to Search Results
Model Number 471172-17
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2023
Event Type  malfunction  
Manufacturer Narrative
A return material authorization (rma) was issued to the customer requesting to have the intuitive device returned.However, isi has not received the product involved with the alleged issue to perform failure analysis.Additional information is being gathered to determine the contribution of the device to the customer reported issue.A follow-up mdr will be submitted if the product is returned (post failure analysis evaluation).
 
Event Description
It was reported that during a da vinci-assisted liver resection surgical procedure, the insulation sheath melted.The procedure was completed with no reported injury.Intuitive surgical inc.(isi) contacted the site and obtained the following additional information regarding this event: the instrument was inspected prior to use and there was no damage or anything out of the ordinary.The thermal damage was first observed intraoperatively during use.According to the surgeon, the thermal damage to the instrument was probably caused by too high energy effect.The instrument did not collide with an energy instrument during the procedure and there was no arcing observed during the procedure.
 
Event Description
Refer to h10/h11 for follow-up information.
 
Manufacturer Narrative
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The maryland bipolar forceps was analyzed.The instrument exhibited charring and localized melting between the grips.This failure is caused in most cases by insulation degradation and charred tissue forming a conductive path.The device has passed the electrical continuity test.The complaint was confirmed by failure analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
MARYLAND BIPOLAR FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18597732
MDR Text Key333982196
Report Number2955842-2024-10716
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874119792
UDI-Public(01)00886874119792(11)230803(10)K12230803
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K214095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number471172-17
Device Catalogue Number471172
Device Lot NumberK12230803 0018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received03/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/03/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-