Catalog Number 1804350-33 |
Device Problems
Difficult to Remove (1528); Device Dislodged or Dislocated (2923)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/06/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
|
Event Description
|
It was reported that before use in the patient, the protective sheath of the xience skypoint des 3.50 x 33 rx us was difficult to remove and the stent separated from the balloon.Another xience was used to complete the procedure.There were no reported adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
A visual inspection was performed on the returned device.The reported stent dislodgement was confirmed.The reported difficult to remove could not be tested due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.A conclusive cause for the reported difficulties could not be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.Na.
|
|