SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD ALARIS EXTENSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 20129E
Device Problem Complete Blockage (1094)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter facility name- (b)(6) specialists.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event Description
It was reported that bd alaris extension set was occluded the following information was received by the initial reporter with the following verbatim: item description: tube iv ext w/ filter 1.2 mi lawson# (b)(4).Mfg# 20129e lot# n/a qty: (b)(4).Po #(b)(4).Description of problem: the pump was reading occlusion, i tried trouble shooting prior to placing a new filter onto the tubing.Once the new filter was placed, the pump no longer read occlusion.
 
Event Description
No additional info.
 
Manufacturer Narrative
The customer reported that the set occluded.One sample model 20129e was returned for investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.The set was attached to a bd primary set and primed with 85% glycerin / 15% water an infusion run was started at a rate of one ml/hr for 24 hrs.There was no observation of fluid blockage.The customer complaint was unable to be replicated.See the attached ¿filtering out the facts: recommendations to optimize performance of in-line filters for parenteral nutrition and intravenous fat emulsion infusions (ivfe)¿ for additional information regarding the use of ivfe with in-line filter administration sets.Although the issue was unable to be replicated and the root cause is unknown bd is looking at opportunities to improve the filter function to alleviate this failure in the future.A device history record review could not be performed because a lot number was not provided by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS EXTENSION SET
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18603605
MDR Text Key335085775
Report Number9616066-2024-00110
Device Sequence Number1
Product Code FPA
UDI-Device Identifier50885403234901
UDI-Public(01)50885403234901
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K790108
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number20129E
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/29/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-