Continuation of d10: product id 37642 lot# serial# (b)(6), implanted: explanted: product type programmer, patient section d information references the main component of the system.Other relevant device(s) are: product id: 37642, serial/lot #: (b)(6), ubd: , udi#: (b)(4), h3: analysis of the device, model # 37642, s/n (b)(6), revealed telemetry board, intermittent failure and cracked case.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
It was reported that the patient programmer (pp) was "very worn out," which included the battery compartment cover coming off and breaking.As a result, the inside of the battery compartment became worn as well, to the point that they could no longer see the serial number.The patient added that sometimes the pp would not power on, even after they put in brand-new aaa batteries.The patient confirmed they were using appropriate aaa batteries.The patient noted that they change the aaa batteries frequently due to the power issue.The patient mentioned that they had an appointment with their healthcare provider (hcp) yesterday and were asked if they ever check their implantable neurostimulator (ins), and the patient told them that they do not use the pp because of the issues they had with it.A replacement pp was sent.No patient symptoms were reported.
|