• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH SAGITTAL SAW ATTACHMENT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH SAGITTAL SAW ATTACHMENT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.021
Device Problems Device Slipped (1584); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2024
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Event Description
It was reported from france that after a surgical procedure the sagittal saw attachment device wheel rotated and came away from the tip head.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device had cosmetic damage, the coupling was worn and did not function.The device also failed pretests for check function of quick saw blade coupling and check oscillation frequency with frequency meter.The assignable root cause was traced to component failure due to faulty parts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAGITTAL SAW ATTACHMENT
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key18606554
MDR Text Key334098833
Report Number8030965-2024-01568
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07611819107182
UDI-Public07611819107182
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K971544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.021
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-