It was reported to abbott, a patient was having problems with their hardware and enough other structural problems that they were no longer getting relief.Surgical intervention was taken where the whole drg system was explanted.Due to scar tissue, the leads broke while trying to remove them from the ipg pocket.A segment of each lead was left in the body.The results of the investigation are inconclusive as the device was not returned for evaluation.Based on the information received, a single definitive root cause for the issue encountered was unable to be conclusively determined.
|
Related manufacturer reference number:1627487-2024-00461 it was reported that during a system explant on (b)(6) 2024 ( reported in manufacturer report: 1627487-2024-00499,1627487-2024-00498), the physician was unable to explant the leads completely due to the leads being scarred and fragmented when trying to remove them from the ipg pocket.The fragmented portions of the leads were left implanted, and the procedure was completed.Surgical intervention may take place to remove the remaining lead fragments at a later date if needed.
|