• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN LINER; PROTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN LINER; PROTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problem Insufficient Information (4580)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Product remains implanted and will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2024-00332; 0001822565-2024-00333; 0001822565-2024-00335.
 
Event Description
It was reported that the patient underwent a hip arthroplasty on an unknown date.Subsequently, the patient was being considered for a revision for an unknown reason; however, it was determined on an unknown date that the patient did not end up needing or having a revision.No additional information available.
 
Event Description
No further event information is available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned, or pictures provided; visual and dimensional evaluation could not be performed.Radiographs were provided and reviewed by a health care professional.A review of the available records identified the following: right total hip arthroplasty with oversized acetabular cup and shallow acetabular inclination angle.Part and lot identification are necessary for review of device history records and compatibility checks, and neither were provided.Additionally, no problem was found with the devices upon further follow up.It was reported that the implant identification was not related to a revision and was just a follow up visit with a new patient.Additionally, the size and placement of the acetabular components are to the surgeon's discretion and does not indicate an implant issue.This complaint could not be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN LINER
Type of Device
PROTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18608577
MDR Text Key334113606
Report Number0001822565-2024-00329
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-