DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-UC-3023-53 |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/19/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) the cs300 intra-aortic balloon pump (iabp) the display had horizontal lines on the bottom as well as blurry spots.There was no patient involvement reported.
|
|
Manufacturer Narrative
|
The service territory manager (stm) evaluated the unit on a later date.The stm successfully completed a function check with testing catheter and trainer without issue upon initially testing unit.In order to solve the issue, the stm replaced display, display grid/casing, and display cable, and successfully obtained a clear display without issues.The coiled cord, when attempted out of the box, produced electrical test fails code 120.The existing coiled cord was left on unit, fully functional, which would have been replaced as a precaution.Preventative maintenance (pm) was completed successfully with measurements and calibration information separately.The unit was calibrated and passed all functional and safety tests per factory specifications.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, g3, g6, h2, h10, h11 corrected fields: d1, d4.
|
|
Event Description
|
N/a.
|
|