MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-UC-3023-53
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) the cs300 intra-aortic balloon pump (iabp) the display had horizontal lines on the bottom as well as blurry spots.There was no patient involvement reported.
 
Manufacturer Narrative
The service territory manager (stm) evaluated the unit on a later date.The stm successfully completed a function check with testing catheter and trainer without issue upon initially testing unit.In order to solve the issue, the stm replaced display, display grid/casing, and display cable, and successfully obtained a clear display without issues.The coiled cord, when attempted out of the box, produced electrical test fails code 120.The existing coiled cord was left on unit, fully functional, which would have been replaced as a precaution.Preventative maintenance (pm) was completed successfully with measurements and calibration information separately.The unit was calibrated and passed all functional and safety tests per factory specifications.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, g3, g6, h2, h10, h11 corrected fields: d1, d4.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18608751
MDR Text Key334514255
Report Number2249723-2024-00375
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108704
UDI-Public10607567108704
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-UC-3023-53
Device Catalogue Number0998-UC-3023-53
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/06/2007
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-