Catalog Number 07826915700 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/09/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The analyzer serial number is: (b)(6).The customer replaced the hemolyzing reagent.The investigation is ongoing.
|
|
Event Description
|
There was an allegation of questionable a1c-2 tina-quant hemoglobin a1c gen.2 results for 1 patient sample on a cobas pure c 303 analytical unit.The customer reported a positive bias on their proficiency results which prompted them to performed a patient comparison.The initial a1c-2 was 6.01%.The sample was repeated on a c11 analyzer and the result was 5.46%.
|
|
Manufacturer Narrative
|
Based on the calibration and qc data, a general reagent issue could be excluded.The customer changing the hemolyzing reagent resolved the issue.
|
|
Search Alerts/Recalls
|