SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE BLOOD/IV FLUID WARMER; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Model Number HL-90 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
B3: date of event and d5.Operator of device is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the power would not turn on.The event occurred before patient use, and there was no harm/adverse event reported.
|
|
Manufacturer Narrative
|
H3.Device evaluated by manufacturer and h6.Evaluation codes: updated.Device evaluation: one device was received.Per visual inspection, cracks were observed on the cover of the circulating water tank.Per functional testing, the main unit did not turn on even if the power switch was turned on.The complaint was confirmed.The root cause was the main pc board assembly failure.It was unknown what caused the failure.The product's history records were reviewed and there were no non-conformance's nor service-related issues that would have resulted in the reported complaint.At the customer's request, the product was returned to the without repair.
|
|