Model Number 471205-17 |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 10/02/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The fenestrated bipolar forceps instrument was analyzed and found the primary failure of broken grip cable to be related to the customer reported complaint.The instrument was found to have a broken grip cable at the proximal end.This causes the grasper not to function properly.Additional observation related to customer reported complaint: the instrument was found to have a loose pitch cable at the distal end.The loose pitch cable at the distal end was caused by the broken pitch cable on the proximal end.In addition, the instrument was found to have charring and/or localized melting on the outer surface of one bipolar yaw pulley, resulting in an exposed electrode.The instrument passed the electrical continuity test.The complaint was confirmed by failure analysis.
|
|
Event Description
|
It was reported that during a da vinci-assisted surgical procedure, the fenestrated bipolar forceps instrument's graspers would not function properly.The procedure was completed with no injury to the patient.
|
|
Manufacturer Narrative
|
The fenestrated bipolar forceps instrument was further evaluated, and it was found to have a broken proximal pitch cable at the proximal end.This causes the grasper not to function properly.Additionally, the instrument was found to have charring and/or localized melting on the outer surface of one bipolar yaw pulley, resulting in an exposed electrode.The instrument passed the electrical continuity test.
|
|
Event Description
|
Refer to h10/h11 for follow-up information.
|
|
Search Alerts/Recalls
|