Model Number 21-7346-24
Device Problem Inaccurate Delivery (2339)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2023
Event Type  malfunction  
Event Description
It was reported that there was an infusion problem while running iv antibiotics.Per reporter the line was changed on (b)(6) but exhibited problems with infusing of antibiotics.On december 13th the cadd pump was changed out to hc cadd pump.Problems persisted and cadd tubing was changed on december 14th - with no further concerns with infusing antibiotic.No adverse patient effects were reported by the customer.
 
Manufacturer Narrative
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
D9: date returned to mfg 3/4/2024.One sample was returned for evaluation.Visual inspection revealed no damage or other defects.Functional testing was performed and no discrepancies were detected.The failure mode was not confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET, ADMIN, CADD, 94", SPIKE, 0.2 FLTR, COILED TUBE, FS, TOTM 15/BX
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan ln n
minneapolis, MN 55442
2247062300
MDR Report Key18610931
MDR Text Key334511370
Report Number9617604-2024-00087
Device Sequence Number1
Product Code FPA
UDI-Device Identifier15019517161953
UDI-Public15019517161953
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K031361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7346-24
Device Catalogue Number21-7346-24
Device Lot Number4396148
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-