Model Number 11500A
Device Problems Fluid/Blood Leak (1250); Gradient Increase (1270)
Patient Problems Chest Pain (1776); Dyspnea (1816); Fatigue (1849); Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Event Date 01/08/2024
Event Type  Injury  
Manufacturer Narrative
The subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
It was reported a patient with a 23mm 11500a aortic valve implanted for three (3) years, seven (7) months, has been placed under evaluation for a valve-in-valve procedure due to stenosis and aortic insufficiency.
 
Event Description
It was reported a patient with a 23mm 11500a aortic valve implanted for three (3) years, seven (7) months, has been placed under evaluation for a re-intervention procedure due to severe aortic stenosis.The patient presented with nyha class iii, shortness of breath, chest pain, and fatigue.No planned intervention currently.
 
Manufacturer Narrative
The device history record (dhr) review was completed and there were two non-conformances found related to this serial number.However, the valve was re-worked to completion, passing all inspections.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction related to a manufacturing deficiency.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The most likely cause is patient factors, including coronary artery disease (cad), history of coronary artery bypass graft (cabg), chronic kidney disease (ckd), and hyperlipidemia (hld).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INSPIRIS RESILIA AORTIC VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key18614639
MDR Text Key334213963
Report Number2015691-2024-00678
Device Sequence Number1
Product Code LWR
UDI-Device Identifier00690103194982
UDI-Public(01)00690103194982(17)241210(11)191212
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number11500A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening;
Patient Age64 YR
Patient SexMale
-
-