E1 initial reporter phone: (b)(6).The product investigation was completed.Device evaluation details: the thermocool® smart touch® sf bi-directional navigation catheter device was returned to biosense webster (bwi) for evaluation.A visual inspection and screening test of the returned device were performed following bwi procedures.Visual analysis of the returned sample revealed a hole and reddish material in the pebax.A magnetic sensor functionality test was performed and the device was recognized on the system; however, error 105 was displayed on the screen due to an open circuit on the tip area.The blood found inside the pebax area may contribute to the magnetic issue and the force issue reported by the customer.A manufacturing record evaluation was performed for the finished device 31104101l number, and no internal action was found during the review.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use (ifu) contain the following recommendations: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number:(b)(4).
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and post procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, the contact force (cf) suddenly increased during ablation.The vector was also out of alignment and zeroing was performed several times but did not resolve the issue.This issue occurred during the 4th ablation.When the physician pulled out the catheter to replace it, they found that blood had accumulated in the spring portion of the catheter.The catheter was wiped, but there was no thrombus.The problem was not resolved by unplugging and plugging nor by replacing the cable.The issue was resolved by replacing the smart touch sf catheter to another new one.No patient consequences were reported.
|