Model Number 71953-01
Device Problem Power Problem (3010)
Patient Problems Hyperglycemia (1905); Neuropathy (1983)
Event Date 12/25/2023
Event Type  Injury  
Event Description
A customer reported being unable to test using the adc device due to a fast-draining battery and as a result, the customer experienced symptoms of "high sugar, neuropathy." customer was unable to self-treat and had contact with a healthcare professional who provided an unspecified iv and an unspecified (dose/type) insulin twice per day as treatment for the diagnosis of hyperglycemia.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history record) for the freestyle libre reader, and kit pack for verification of the correct cable and adapter were reviewed and the dhrs showed the freestyle libre reader passed all tests prior to release and the correct cable and adapter were part of the kit pack.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18618944
MDR Text Key334255650
Report Number2954323-2024-03899
Device Sequence Number1
Product Code QLG
UDI-Device Identifier00357599803001
UDI-Public00357599803001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71953-01
Was Device Available for Evaluation? No
Date Manufacturer Received01/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/21/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-