Model Number A219 |
Device Problems
Premature Discharge of Battery (1057); Incorrect, Inadequate or Imprecise Result or Readings (1535); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/27/2023 |
Event Type
Injury
|
Event Description
|
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) displayed a bd message alert indicative of a potential premature battery depletion (pbd).The device battery decreased to 15%.The device was explanted and replaced.No additional information is available at the moment.No additional adverse patient effects were reported.Additional information was obtained, the device was explanted and replaced.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, this s-icd was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of a compromised capacitor.This resulted in the observed premature battery depletion.It was determined that the capacitor was compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of emblem devices that have a potential of exhibiting this behavior.This particular device is included in the emblem s-icd accelerated depletion advisory population.
|
|
Event Description
|
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) displayed a bd message alert indicative of a potential premature battery depletion (pbd).The device battery decreased to 15%.The device was explanted and replaced.No additional information is available at the moment.No additional adverse patient effects were reported.Additional information was obtained, the device was explanted and replaced.
|
|
Search Alerts/Recalls
|