Model Number PM2152 |
Device Problem
No Pacing (3268)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported, during the initial implant procedure after connecting the leads to the device, there was no pacing observed.The connection was check and pacing was still unable to be observed.The device was exchanged to resolve the event.The patient was stable.
|
|
Manufacturer Narrative
|
Further information was requested but not received.
|
|
Manufacturer Narrative
|
The reported event of no pacing was not confirmed.The device was still in original shipping conditions upon receipt with battery voltage near beginning of life level.There were no measurements made on the device in the field.Electrical and mechanical evaluation performed under nominal settings indicated normal functionality.Further analysis of output parameters found normal pacing with all parameters within normal range.Additionally, visual inspection of the header and connector detected no contamination or foreign material that could contribute to the reported event.Longevity assessment was performed, and the device was in normal range of operation with appropriate remaining longevity.
|
|