Model Number PM2152
Device Problem No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported, during the initial implant procedure after connecting the leads to the device, there was no pacing observed.The connection was check and pacing was still unable to be observed.The device was exchanged to resolve the event.The patient was stable.
 
Manufacturer Narrative
Further information was requested but not received.
 
Manufacturer Narrative
The reported event of no pacing was not confirmed.The device was still in original shipping conditions upon receipt with battery voltage near beginning of life level.There were no measurements made on the device in the field.Electrical and mechanical evaluation performed under nominal settings indicated normal functionality.Further analysis of output parameters found normal pacing with all parameters within normal range.Additionally, visual inspection of the header and connector detected no contamination or foreign material that could contribute to the reported event.Longevity assessment was performed, and the device was in normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY¿ CORE DR, JPD
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18619054
MDR Text Key334255749
Report Number2017865-2024-03453
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeBU
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPM2152
Device Lot NumberP000178458
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-