It was reported that a device entrapment occurred.The 100% stenosed target lesion was located in the mildly tortuous and calcified posterior tibial artery.A 1.75mm rotapro and a rotawire were selected for use.During the procedure, when advancing the burr through the lesion, the burr was not spinning, and the wire was flopping on the back table indicating that the burr was stuck on the wire.The burr and wire were removed together as one unit.There seemed to be an intimal tissue wrapped around the catheter tip when removed, that was suspected to be the cause of the device became stall and stuck.The procedure was completed with another of the same device.There were no patient complications reported post procedure.
|