• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM; SUTURE MEDIATED CLOSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM; SUTURE MEDIATED CLOSURE Back to Search Results
Catalog Number 12773-03
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2024
Event Type  Injury  
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event Description
It was reported that suture placement in the right common femoral vein was attempted with a prostyle device using the pre-close technique via a 6f sheath hole prior to a thrombectomy interventional procedure.Reportedly, the foot was unable to be opened due to resistance.The sutures of two new prostyle devices were successfully pre-placed.The sheath was upsized to a 24f and the thrombectomy procedure was completed.Hemostasis was achieved with the pre-placed prostyle sutures.There was no reported adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
 
Manufacturer Narrative
The device was returned for analysis.The reported foot deployment issue was not confirmed.A review of the manufacturing records revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties; however, the treatment appears to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM
Type of Device
SUTURE MEDIATED CLOSURE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18619942
MDR Text Key334256218
Report Number2024168-2024-01294
Device Sequence Number1
Product Code MGB
UDI-Device Identifier08717648235184
UDI-Public08717648235184
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number12773-03
Device Lot Number3091642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
Patient SexMale
Patient Weight132 KG
-
-