Model Number ZM-531PA
Device Problems Erratic or Intermittent Display (1182); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2024
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer (bme) reported that the transmitter was experiencing sporadic ecg waveform output.The unit would also read the heart rate (hr) low, then spike high and alarm asystole.The bme tested the transmitter on a simulator and received the same results.There was no physical damage or fluid intrusion.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.The following field contains no information (ni), as attempts to obtain the information were made, but not provided: additional device information: d10 concomitant medical device: the following devices were used in conjunction with the transmitter: cns: model #: ni, serial (b)(6), device manufacturer data: ni, unique identifier (b)(4), returned to nihon kohden: na.Org: model #: org-9110a, serial (b)(6), device manufacturer data: 25/02/2021, unique identifier (b)(4), returned to nihon kohden: na.
 
Event Description
The biomedical engineer (bme) reported that the transmitter was experiencing sporadic ecg waveform output.The unit would also read the heart rate (hr) low, then spike high and alarm asystole.No patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-531PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key18622711
MDR Text Key334850789
Report Number8030229-2024-04106
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921115107
UDI-Public04931921115107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-531PA
Device Catalogue NumberZM-531PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/19/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CNS; ORG
Patient Age69 YR
Patient SexMale
Patient Weight70 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-