Product complaint (b)(4).Investigation summary it was reported that the femoral impactor found to be damaged from impacting of implants after it came out of the wash.Replacement needed.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned device found that the attune system impactor was severely deformed from the distal end.The observed condition of the device was consistent with a component failure that may have been caused by exposure to unintended forces, like impacting of the device before is fully seated.The overall complaint was confirmed as the observed condition of the attune system impactor would contribute to the alleged device issue.Based on the investigation findings, the potential cause is traced to unintended user error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|