• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH TARGET DEVICE GAMMA3®; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TARGET DEVICE GAMMA3®; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 13200111
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2024
Event Type  malfunction  
Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event Description
As reported by the rep: "the surgeon approached me following a short gamma3 nail.He was distally targeting the screws and aimed for the static hole but the screw was placed in the dynamic position and in his words appeared to catch on the edge of the hole.The procedure was completed successfully as the screw went into the dynamic hole position.".
 
Event Description
As reported by the rep: "the surgeon approached me following a short gamma3 nail.He was distally targeting the screws and aimed for the static hole but the screw was placed in the dynamic position and in his words appeared to catch on the edge of the hole.The procedure was completed successfully as the screw went into the dynamic hole position.".
 
Manufacturer Narrative
Please note correction to section d9/h3 the device was lost by the customer.The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.H3 other text : device lost by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET DEVICE GAMMA3®
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18623720
MDR Text Key334495572
Report Number0009610622-2024-00051
Device Sequence Number1
Product Code HSB
UDI-Device Identifier04546540716774
UDI-Public04546540716774
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K123401
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number13200111
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/08/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-