• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST; FENESTRATED BIPOLAR FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST; FENESTRATED BIPOLAR FORCEPS Back to Search Results
Model Number 471205-17
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2024
Event Type  malfunction  
Event Description
It was reported that during a da vinci-assisted surgical procedure, the fenestrated bipolar forceps instrument would not come out easily during exchange.It was indicated that the surgeon scrubbed back in to retrieve the instrument with a little force and inspection upon removal identified thermal damage on the instrument plastic tip piece.The procedure was completed with no reported injury.
 
Manufacturer Narrative
Isi has received the fenestrated bipolar forceps instrument; however, failure analysis is still ongoing.Additional information is being gathered to determine the contribution of the device to the customer reported issue.A follow-up mdr will be submitted post failure analysis evaluation.
 
Manufacturer Narrative
Intuitive surgical, inc.(isi) received the fenestrated bipolar forceps instrument to perform failure analysis.The fenestrated bipolar forceps instrument was analyzed and found to have thermal damage on the yaw pulley.The instrument was placed and driven on an in-house system.The instrument passed the recognition and engagement tests.The instrument attached to and removed without any issues on three of three attempts.The instrument grips were manually manipulated, and the instrument passed an electric continuity test on three of three attempts.The instrument delivered very low energy on three of three attempts.
 
Event Description
Refer to h10/h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
FENESTRATED BIPOLAR FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18624074
MDR Text Key334334098
Report Number2955842-2024-10899
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874119808
UDI-Public(01)00886874119808(10)K12220314
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K214095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number471205-17
Device Catalogue Number471205
Device Lot NumberK12220314 0267
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2024
Date Manufacturer Received02/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-