• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE CENTRELLA MED-SURG; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE CENTRELLA MED-SURG; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P7900B100010
Device Problem Intermittent Continuity (1121)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2024
Event Type  malfunction  
Event Description
Baxter received a report from a baxter technician stating the bed was moving by itself.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
 
Manufacturer Narrative
The baxter technician found the caregiver board needed to be replaced.Per the hillrom service manual, it is necessary for the centrella¿ bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Do the function checks as applicable for your bed.If the bed passes all of the checks for its configuration, do the necessary administrative tasks, and prepare the bed to be put into service.If the bed does not pass all of the checks, repair or replace the part as applicable.Do not put the bed into service until it passes all of the checks.Warning¿report to bed authorized maintenance persons any unusual sounds, burning odors, or movement deviations observed in the controls, motors, or limit switch functions.Check the cable connection between the mcb and bcb.Replace or connect the cable as necessary.Replace the bcb.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the caregiver board to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRELLA MED-SURG
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
estephania cardenas
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key18625272
MDR Text Key334388925
Report Number1824206-2024-00160
Device Sequence Number1
Product Code FNL
UDI-Device Identifier00887761985162
UDI-Public010088776198516211190802
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP7900B100010
Was the Report Sent to FDA? No
Date Manufacturer Received01/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-