As reported, proper hemostasis was not achieved after using a 5f mynxgrip vascular closure device (vcd).The device was removed from the patient¿s body.Hemostasis was achieved by manual compression for less than 30 minutes.There was no reported patient injury.The diagnostic procedure used a retrograde approach.The femoral artery¿s suitability was verified on angiography, including the insertion angle (30-45 degrees) of the vascular sheath introducer.The vessel diameter was verified to be greater than or equal to 5mm in diameter.There was little vessel tortuosity.There was no presence of pvd / calcium in the vicinity of the puncture site.The sheath was not kinked/bent upon removal.There was no visible bend/damage in distal end of the balloon shaft after removal.There was no excess force applied during insertion.The device was stored, prepared and used according to the instructions for use (ifu).The user is trained to the device.The device was opened in a sterile field.Other procedural details were requested but are unknown, unavailable, not answered, or not applicable.The device will be returned for evaluation.
|