MEDTRONIC HEART VALVES DIVISION EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number D-EVOLUTFX-34 |
Device Problems
Failure to Advance (2524); Deformation Due to Compressive Stress (2889)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/19/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: upon receipt at medtronic¿s quality laboratory, the delivery catheter system (dcs) was received without a valve loaded within the capsule.A guidewire was stuck within the dcs wire lumen.Attempts to removal the guidewire were unsuccessful.The device was received with the capsule partially opened.The capsule appeared intact with no evidence of damage.The exposed guidewire distal to the nosecone was kinked and the guidewire outer threading was detached from the inner wire at the kink site.The handle appeared intact.The device was returned with the end cap/screw gear snap fit connected.The exposed guidewire proximal to the wire lumen flush port was kinked and the guidewire outer threading was detached from the inner wire at the kink site.The inner member shaft and spindle hub appeared intact with no evidence of damage.The deployment knob appeared to retract and advance the capsule.The trigger moved to fully advanced and retracted positions and locked in place when released.The tip-retrieval mechanism appeared intact.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during implant of this transcatheter bioprosthetic valve, the patient had challenging anatomy: tortuous legs and horizontal annulus.There was no reported difficulty in advancing the delivery catheter system (dcs).The valve was deployed successfully. the operator was pacing on the guidewire. when the implanting team attempted to remove the system, the dcs was "parked" in the descending aorta and the guidewire could not be retrieved.The team tried using excessive force, however the non-medtronic (boston safari) guidewire was completely jammed inside the dcs.The team had to withdraw the system with the wire together, and close the leg directly without a sheath.Of note, there was no visible damage to the guidewire, however a slight bend at entry of the dcs was reported.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Updated h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|