U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801188
Device Problem Calibration Problem (2890)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that the electronic patient gas system (epgs) would not calibrate.There were no reported adverse consequences to the patient.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The field service representative (fsr) verified the reported complaint via the data log where multiple oxygen (o2) sensor and blender disagree messages were found.The fsr replaced the o2 sensor.The unit operated to the manufacturer's specifications.
 
Manufacturer Narrative
Updated block: h9.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key18630735
MDR Text Key334530469
Report Number1828100-2024-00044
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000588
UDI-Public(01)00886799000588(11)171128
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-