Catalog Number 5C4471 |
Device Problem
Unintended Electrical Shock (4018)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/04/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported the patient felt ¿a jolt¿ in the hand from a homechoice automated pd system when turning on the machine at the beginning of therapy.At the time of the event, the patient was touching ¿the little iron¿ where they placed the cassette set.No signs of burns or other injuries observed on the patient.The patient was standing without shoes.The machine was plugged into a grounded outlet with a "cheater adapter" and no damage noted to the three-prong plug or power cord.There was no moisture on or around the machine and no leaks were detected.There was no report of medical intervention associated with this event.No further information was available at the time of this report.
|
|
Manufacturer Narrative
|
Additional information was added to d9, h3, h6 and h10.H10: the device was received for evaluation.A visual inspection and functional testing were performed with no issues noted related to the reported condition.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|