The meter and test strips were provided for investigation where they were tested.Testing results: (qc range level 1: 30-60 mg/dl; level 2: 261-353 mg/dl).Testing results with the returned meter, retention strips, and retention controls: level 1: 47 mg/dl, 46 mg/dl, 46 mg/dl.Level 2: 317 mg/dl, 317 mg/dl, 314 mg/dl.All results are within acceptable range.No information was provided in the complaint case that would point to a cause for the result discrepancy.Testing with the returned strips was also acceptable.The customer's allegation of questionable results was not observed during testing of returned product, all testing with the returned strips produced acceptable results.This complaint is not verified due to the questionable results allegation not being observed during the investigation of the returned product.Strips batch record review didn't identify any non-conformances issued for this lot related to the customer's allegation.Additionally, retention testing is performed every 3 months an no failures have been observed with this lot.Monitoring for this incident has not exceeded any thresholds or identified any field safety corrective actions.No issues with the meter display were noted during the investigation.The investigation did not identify a product problem.The cause of the event could not be determined.
|