Model Number UNKNOWN ENDO STITCH SULU |
Device Problem
Break (1069)
|
Patient Problems
Foreign Body In Patient (2687); Unintended Radiation Exposure (4565); Insufficient Information (4580)
|
Event Date 01/16/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
D10 concomitant products: 173016, 173016 endo stitch instrument (lot#j3a0120ey) unknown endo st, unknown endo stitch sulu (lot#unknown) unknown estitch, unknown endo stitch instrument (lot#unknown) unknown endo st, unknown endo stitch sulu (lot#unknown) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a laparoscopic paraoesophageal diaphragmatic hernia repair-abdomen fundoplication, the first needle broke off, leaving a piece of it on the handpiece and the tip inside the patient.The surgeon ordered an x-ray and searched the field with no luck.The second device, the needle, is lodged in the unit, but it was bent.The third device was used; the whole needle was left in the patient, who was able to retrieve it.
|
|
Manufacturer Narrative
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the needle was returned loaded in an instrument and was removed for further inspection.The needle was broken at the attachment site.No suture was present with needle.It was reported that the needle was broken.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.Bent or broken needle may occur when the needle was not loaded properly or when the needle was forced into an obstacle.This condition may also occur if excess pressure was exerted on the needle or the attached suture while the jaws were in the open position.In these situations, the needle may flex and ultimately break.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: ensure that the toggle levers are fully retracted prior to opening the jaws of the device.The reloading button should never be pressed when the instrument is in the body cavity, as this will release the needle.Inspect the application site to ensure hemostasis.Place additional sutures or use electrocautery if necessary to complete hemostasis.Endoscopic procedures should be performed only by physicians having adequate training and familiarity with endoscopic techniques.A thorough understanding of the operating principles, risks versus benefits, and the hazards involved in utilizing an endoscopic approach is necessary to avoid possible injury to the user and/or patient.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|