• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0702-XTW
Device Problem Incomplete Coaptation (2507)
Patient Problem Mitral Valve Insufficiency/ Regurgitation (4451)
Event Date 01/17/2024
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information, the cause of the reported slda was unable to be determined.The reported recurrent mr was a cascading event of the reported slda.The reported patient effect of mitral regurgitation, as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.The reported unexpected medical intervention and hospitalization were results of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event Description
It was reported that on (b)(6) 2024, a mitraclip procedure was performed to mixed degenerative mitral regurgitation (mr) with a grade of 3-4.One clip was implanted, reducing mr to a grade of 1.On 17jan2024, echocardiography showed the implanted clip had detached from the posterior leaflet and remained attached to the anterior leaflet (single leaflet device attachment/slda), causing mr to increase to 3.On (b)(6) 2024, a second mitraclip procedure was performed in an attempt to stabilize the slda.However, the mitraclip was unable to be implanted and mr remained at a grade of 3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18636433
MDR Text Key334491041
Report Number2135147-2024-00479
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0702-XTW
Device Lot Number30620R1050
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-