• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 DEPUY CMW 2G 20G; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 DEPUY CMW 2G 20G; BONE CEMENT : BONE CEMENT Back to Search Results
Catalog Number 3325020
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2023
Event Type  malfunction  
Manufacturer Narrative
Product complaint (b)(4).Investigation summary damaged product under sealed box.Please can you send me the replacement for this lot no 4076141 exp 31/01/2026 mfg- 02-06-2023.Dsi case (b)(4).The product was not returned to depuy synthes, however photos were provided for review.(b)(4).The photo investigation revealed the liquid monomer solidified inside the original sealed package, suggesting the ampule broke.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed for the [depuy cmw 2g 20g].The root cause of the complaint condition cannot be traced to a manufacturing facility.However, once the product leaves depuy synthes control, it is unknown what environment conditions the packaged products are exposed to during that time.Therefore the suspected cause is traced to transport/storage outside of depuy synthes control.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that device within the sealed packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPUY CMW 2G 20G
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND UC
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18637370
MDR Text Key334498035
Report Number1818910-2024-02291
Device Sequence Number1
Product Code LOD
Combination Product (y/n)N
Reporter Country CodeNZ
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3325020
Device Lot Number4076141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-