The event log was reviewed and confirms that the pump alarmed system error during an infusion.This is seen on lines 557, 558, 561 and 562.There are multiple subcodes for the system error.24 indicates motor is fast, motor delay issue.44 indicates motor open, motor delay issue.Event 557: log_event_alarm time: 08:50:47 alarm code: 02 sub code: 24 alarm status: log_alarm_cause_ack_alarm eventbytes: 05 50 47 02 24 00 31 f3.Event 558: log_event_alarm time: 08:51:07 alarm code: 02 sub code: 24 alarm status: log_alarm_cause_exit_to_prog eventbytes: 05 51 07 02 24 00 32 b5.Event 561: log_event_alarm time: 08:52:09 alarm code: 02 sub code: 44 alarm status: log_alarm_cause_ack_alarm eventbytes: 05 52 09 02 44 00 31 d7.Event 562: log_event_alarm time: 08:52:15 alarm code: 02 sub code: 44 alarm status: log_alarm_cause_exit_to_prog eventbytes: 05 52 15 02 44 00 32 e4.A performance test was done on the pump where an 100 ml infusion was started, and the pump alarmed system error 3 ml's into the infusion.The reported issue is confirmed.The pump does not meet passing criteria.
|