Model Number HYF-1T
Device Problem Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was observed that during the device evaluation, the hysterofiberscopes light guide cover glass had a foreign body.There were no reports of patient involvement.
 
Manufacturer Narrative
The device was returned and evaluated.In addition to the malfunction documented in b5, the additional evaluation findings are as follows: the light guide (lg) rod and control unit both have a burr, the lg connector, universal cord, grip, diopter ring, and the eye piece all have a scratch, the grip, control unit, and lg lens all have discoloration, control unit has a crack, due to wear of angle wire; the bending angle in the up direction does not meet the standard value, and the image guide bundle has significant breakages.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material could not be identified nor the root cause of it.Per additional follow up from the customer, the device was cleaned, disinfected, and sterilized the product before it sent in for repair.According to the customer, it is not known when the foreign material adhered to the nozzle.The event can be detected/prevented in accordance with the following instructions for use: chapter 7 care and storage: 7.1 application of the method of cleaning and disinfection to each product.7.2 condition of cleaning and disinfection.7.3 request about cleaning and disinfection.7.4 cleaning and disinfection.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES HYSTEROFIBERSCOPE
Type of Device
HYSTEROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18643628
MDR Text Key335823093
Report Number3002808148-2024-01099
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K891451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHYF-1T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2024
Was the Report Sent to FDA? No
Date Manufacturer Received02/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-