SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Model Number HL-90 |
Device Problems
Insufficient Heating (1287); Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/26/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
D5.Operator of device and e1: name is unknown, no information has been provided to date.E1 phone number: (b)(6).Device evaluation: product was not returned, and no photographic evidence provided to aid in this investigation.As a result, product evaluation and problem confirmation cannot be performed.The exact cause could not be determined; however, based on the reported problem, the most probable cause is a faulty main board.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
|
|
Event Description
|
It was reported that the device could not heat the liquid that needed to be heated.After testing the equipment, it was found that the sensor on the main board of the machine was faulty, causing the machine to be unable to heat properly and then stopped using it.The faulty machine was repaired by the hospital.No patient involvement was reported.
|
|