Model Number HL-90
Device Problems Insufficient Heating (1287); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/26/2023
Event Type  malfunction  
Manufacturer Narrative
D5.Operator of device and e1: name is unknown, no information has been provided to date.E1 phone number: (b)(6).Device evaluation: product was not returned, and no photographic evidence provided to aid in this investigation.As a result, product evaluation and problem confirmation cannot be performed.The exact cause could not be determined; however, based on the reported problem, the most probable cause is a faulty main board.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Event Description
It was reported that the device could not heat the liquid that needed to be heated.After testing the equipment, it was found that the sensor on the main board of the machine was faulty, causing the machine to be unable to heat properly and then stopped using it.The faulty machine was repaired by the hospital.No patient involvement was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18646269
MDR Text Key334591810
Report Number3012307300-2024-00551
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberHL-90-CN-230
Was Device Available for Evaluation? No
Date Manufacturer Received01/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/17/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse