A lead extraction procedure commenced to remove a right atrial (ra), a right ventricular (rv) and a left ventricular (lv) lead due to non-function.Spectranetics lld ez lead locking devices (lld ezs) were inserted into each lead to provide traction.Beginning with a spectranetics 14f glidelight laser sheath with a medium visisheath dilator sheath on the ra lead, progress stalled due to lead on lead binding.Moving to the lv lead with the same glidelight, the lead was successfully removed.Moving again to the ra lead using the glidelight and visisheath, the remaining lead on lead binding was resolved in the innominate and superior vena cava (svc) regions.While the glidelight and visisheath were still located in the innominate/svc region, the ra lead came free and was removed, along with the glidelight and visisheath; however, the patient''s blood pressure dropped.Rescue efforts began, including rescue balloon and sternotomy.A right atrial appendage (raa) perforation was discovered, clamped, and repaired.Once the patient stabilized, it was decided to abandon extraction of the rv lead.Attempts were made to unlock the lld ez from the rv lead, but were unsuccessful; therefore, the rv lead/lld were cut and capped and remained in the patient (mdr #3007284006-2024-00031).The patient survived the procedure.This report captures the lld ez providing traction within the ra lead when the perforation occurred, requiring intervention.There was no alleged malfunction of the lld ez within the ra lead during the procedure.
|
D4): device lot number, expiration date unk.Partial udi populated.H3): the device was discarded, thus no investigation could be completed.H6): cardiac perforation is a known risk of complication with use of the lld.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|