The device was returned and evaluated, and there was found to be corrosion of the scope connector and the plug unit.Additional findings include the following: leak from biopsy channel, angulation, play in the control know movement, deep dents in the distal end plastic cover, chips in the objective lens, light guide lens, bending section cover glue, buckle near distal end of insertion tube, fluid wet dry after aeration in the control body, light guide tube, scope connector had fluid, wet, dry after aeration, wet detection sticker activated.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|